Omega OM-SQ2010 Portable Data Logger with components, software/installation guide and protective hard carrying case. Unit was purchased New in 2016 and used only briefly for a short project that was cancelled. Included also is Aprox. 200 – 250 ft of Omega Cabling for use with Omega OM-SQ2010 Data Logger and is compatible with other devices as well. Cable is insulated 2x with fiberglass and has a deviation of 0.5 degrees F (Located in Pittsburgh, PA)Manufacturer Specs for the Omega OM-SQ2010 Portable Data Logger:
No. of Analogue Channels:
8 single ended or 4 differential inputs. The OM-SQ2010 data logger has a single analogue to digital converter (A/D) which corresponds to inputs on blocks A and B. Each connection block will accept up to 2 differential inputs or up to 4 single ended inputs (it is not possible to mix single ended and differential inputs on a block).
Analogue Input Connections:
Detachable srew terminal blocks
Channel Expansion:
No
Universal Input:
Yes
Voltage Ranges (Differential and Single Ended):
-6 to 25, -0.6 to 2.4, ±0.3V, -0.15 to 0.15, -0.075 to 0.075,
-6 to 12, -6 to 6, -3 to 3, -0.6 to 1.2, -0.6 to 0.6
Common Mode:
25 V
Current Ranges, Differential (Requires External 10 Ω Shunt):
4 to 20 mA, ±30 mA
Thermocouple Ranges (Differential and Single Ended):
Type J:
-200 to 1200°C (-328 to 2192°F)
Type K:
-200 to 1372°C (-328 to 2502°F)
Type T:
-200 to 400°C (-328 to 752°F)
Type N:
-200 to 1300°C (-328 to 2372°F)
Type R:
-50 to 1768°C (-58 to 3214°F)
Type S:
-50 to 1768°C (-58 to 3214°F)
Resistance Ranges (All 2 Wire):
0 to 1250 Ω, 0 to 5000 Ω 0 to 20,000 Ω, 0 to 300,000 Ω
Thermistor Ranges:
U & UU-Type:
-50 to 150°C (-58 to 302°F),
Y-Type:
-50 to 150°C (-58 to 302°F)
S-Type:
-30 to 150°C (-22 to 302°F)
User-Defined Thermistor:
Enter Steinhart-Hart coefficients or RT pairs
Pt100/1000 (2-wire):
-200 to 850°C (-328 to 1562°F)
A/D Resolution:
24 bit
Accuracy:
See table on pdf data sheet
Internal Reference Temperature:
-50 to 150°C (-58 to 302°F)
Pulse Count Ranges:
0 to 100 Hz (1 input); 0 to 64 kHz (1 input); 0 to 16,000,000 count
Digital State/Event Ranges:
8 state inputs or 1 x 8 bit binary
Digital/Alarm Outputs:
2 open drain FETs, 18V, 0.1A
Digital I/O Connections:
DB25F connector
Clock Resolution/Accuracy:
1s/10 ppm Normal Mode - each input sampled at a max rate of 1 rdg per second
Double-Speed (Mains Reject Off) - One input can be sampled at 10 rdgs per second and all others are sampled at a max rate of 1 rdg per second
No of Intervals:
4
Data Scaling:
Included in standard OM-SQ software
Data Statistics:
Calculated within OM-SQ-SOFT-PLUS software
Calculated Channels:
Up to 16
Memory Internal:
16 M (1 to 14 million readings)
Display/Keypad:
128 x 64 dot graphical display, 4 button keypad
Power:
2 C cells internal (included), or external 8 to 28 Vdc via AC adaptor and USB when plugged in
Battery Life:
Up to 5 days with continuous usage while sampling all channels once per second
Sensor Power Output:
5 V at 50 mA, external 8 to 28V at 100 mA (when connected)
Networking:
Via RS-232 to Ethernet adaptor (Model No. OM-SQ-NET-ADAP)
Modem Support :
Via RS-232 modem (GSM modem kit Model No. OM-SQ-GSM-KIT)
PC Setup:
Complete data logger set up possible via OM-SQ software;
Compatible with Windows XP SP3/Vista/7 and 8 (32-bit and 64-bit)
Front Panel Setup:
Via 4 integral keys. All essential functionality available via key pad e.g. channel configuration, start/stop logging etc. Other advanced functions e.g. calculated channels and channel descriptions are available via connection to a PC running OM-SQ data logger software
Stored Setups:
6
Operating Temperature:
-20 to 65°C
Dimensions:
135 D x 175 W x 55 mm H
Weight:
0.7 kg
Enclosure: